Sper-scientific 840022 Light Meter Advanced Manual de usuario Pagina 5

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 8
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 4
- 5 -
To return to recording mode, press the HOLD (4) button. "REC" is
displayed without "Max" or "Min".
To erase the recorded Max/Min values and exit the recording mode,
press the REC MAX/MIN (5) button for at least 2 seconds.
3-E.
Notes and Precautions
Avoid range overload.
When using PEAK HOLD (3-C) or REC MAX/MIN (3-D) your
measurements must be within the selected range, if not, the out of
range indicator "- - - -" will be displayed.
Cover the PHOTO SENSOR (16) with the sensor cap when the meter
is not in use.
Fluctuations in the reading are often due to shadows or a weak bat-
tery. Also, ambient temperatures and drafts may affect the reading.
Do not store in areas of high temperature and/or humidity.
Remove the battery for long-term storage.
Do not adjust the calibration trimmer beneath the RS232 Cover (6).
4. ZERO ADJUSTMENT AND CALIBRATION
With the sensor cover securely in place and the meter turned on, use
the RANGE (9) button to select the 40.00 Lux range (digits appear).
Press the ZERO (11) button and the "0.00
LUX" should be displayed.
To maintain accuracy, annual laboratory calibration is recommended
and available from Sper Scientific.
5. BATTERY REPLACEMENT
Replace the battery when the low battery icon is displayed.
In-spec measurements may be made for several hours after the low
battery indicator appears.
Remove the BATTERY COVER (14) screw and slide the cover off.
Replace the battery with a fresh 9V battery (alkaline or heavy duty).
6. RS232 PC SERIAL INTERFACE
Use a small screwdriver to gently lift the RS232 COVER (6) up to expose
the RS232 OUTPUT (8) 3.5 mm terminal. The signal output is a 16-digit
data stream that can be adapted to user-defined applications. A RS232
lead with the following connection is required to link the instrument with the
PC serial interface.
Meter (3.5 mm jack plug) PC (9W ‘D” Connector)
Center Pin .................................. Pin 4 Pin 2 2.2 K
Ground/shield ............................ Pin 2 Pin 5 resistor
Vista de pagina 4
1 2 3 4 5 6 7 8

Comentarios a estos manuales

Sin comentarios