Sper-scientific 850055 pH Meter - Advanced Manual de usuario Pagina 27

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 26
27
MAINTENANCE
The following actions will keep the probe in good working
condition.
 Always keep the pH glass bulb wet by using the plastic
bottle to protect and store the probe. You can also store
in a KCl solution. Never use distilled or de-ionized water
for storage.
 Always rinse pH probe in de-ionized water before using.
 Never touch or rub the glass bulb tip.
 This probe is designed with a fiber junction. To prolong
the life of the probe, it is recommended to clean the
probe monthly by immersing in a cleaning solution for a
minimum of 30 minutes. After cleaning, rinse with tap
water and recalibrate with the meter.
 To further prolong the life of the probe, extend the fiber
junction and remove the dirt portion. The extendable
fiber reference junction is used to eliminate the reading
errors from a clogged junction.
Vista de pagina 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 39 40

Comentarios a estos manuales

Sin comentarios