3-D. CALIBRATION
Use Sper Scientic Acoustical Calibrator
850016 (or equivalent) to calibrate the meter.
Perform the calibration at temperatures
between 58~78°F (15~25°C). The calibrator
and meter should be at the same temperature.
Remove the WINDSCREEN (1) and
turn on the meter by pressing the
POWER (9) button.
Turn the acoustical calibrator on to 94.0 dB
and place it onto the MICROPHONE (2).
With the meter in FAST response mode
and A (dBA) weighting, press and hold the
A/C (4) button.
Without releasing the A/C (4) button,
press and hold the HOLD (5) button.
The display will go blank.
Release both buttons and “94.0” ± 0.2
is displayed.
The calibration process can be repeated
until the meter reads the desired value.
This completes the calibration process.
4. AUTO POWER OFF
If no buttons are pushed for about 3 minutes,
the meter will automatically turn off.