2 POINT ACOUSTICAL CALIBRATOR
850016
INSTRUCTION MANUAL
SPER
SCIENTIFIC
Vista de pagina 0
1 2 3 4 5 6

Indice de contenidos

Pagina 1 - SCIENTIFIC

2 POINT ACOUSTICAL CALIBRATOR 850016 INSTRUCTION MANUAL SPER SCIENTIFIC

Pagina 2

INTRODUCTION Calibration is suggested by OSHA to ensure good sampling practices. The 850016 Sper Scientific 2-Point Acoustical Calibrator is designed

Pagina 3

OPERATING PROCEDURE Note: For accurate calibration, background noise should be less than 10 dB. • Determine the sound meter's microphone size.

Pagina 4

BATTERY INSTALLATION AND REPLACEMENT • Replace the battery if the Power Indicator (3) does NOT light when 94dB/114dB/Off Slide Switch (4) is in the 94

Pagina 5

SPECIFICATIONS • Standard: ANSI S1.40-1984 and IEC942 1988 Class 2 • Output Sound Level: 94 dB and 114 dB re 20 uPA under reference • Output Frequency

Pagina 6

WARRANTY Sper Scientific warrants this product against defects in materials and workmanship for a period of 5 years from the date of purchase, and ag