Sper-scientific 850016 Acoustical Calibrator - 2 Point Manual de usuario Pagina 3

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 6
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 2
OPERATING PROCEDURE
Note: For accurate calibration, background noise should
be less than 10 dB.
Determine the sound meter's microphone size.
Use the 1/2" Microphone Adapter (2) when
calibrating a 1/2" diameter microphone.
When calibrating a sound meter with a 1"
microphone, pull off the 1/2" Microphone
Adapter (2) to use the 1" Microphone Insertion
Socket (1).
Fully insert the sound meter's microphone into the
insertion socket.
Use the 94dB/114dB/Off Slide Switch (4) to turn
the calibrator on and select either 94 dB or 114 dB.
Be sure the sound meter is set to an appropriate
scale to measure a matching sound of 94 dB or
114 db.
Adjust the sound meter according to it's calibration
instructions.
Unfold for Battery Replacement,
Specifications and Warranty information.
Vista de pagina 2
1 2 3 4 5 6

Comentarios a estos manuales

Sin comentarios