Free and TotalCHLORINE METERModel : CL-2006Your purchase of this Free and Total CHLORINE METER marks astep forward for you into the field of precision
6)Wait one minute approximately, then insert the " Testing bottle " to the bottom of " Container " ( 3-2, Fig. 1 ) completely.Bef
4-4 Data HoldDuring the measurement, press the " Hold Button " ( 3-4,Fig. 1 ) once will hold the measured value & the LCD willdisplay a
5. CALIBRATION PROCEDURE1)The meter can be calibrated under following calibrationpoints :Free chlorine *Zero *1.00 ppm orTotal chlorine *Zero *1.0
4) Zero chlorine calibration * Insert the " Zero Chlorine standard solution into the " Container " ( 3-2 , Fig. 1) and envelope in th
* Press " CAL Button " ( 3-5, Fig. 1 ) once, the Display will show following text with flashing ( approx.10 seconds ). flashingCAL0.0 ppmTh
* Insert the " 1.0 ppm standard solution into the " Container " ( 3-2 , Fig. 1) and envelope in the " Cover " ( 3-1, Fig.
6. BATTERY REPLACEMENT1)When the left corner of LCD display show " ", itis necessary to replace the battery. However, in-spec.mea
TABLE OF CONTENTS1. FEATURES...12. SPECIFICATIONS...
1. FEATURES* The meter measures the Free and Total chlorine ( CL )in the 0.00 to 3.50 ppm ( mg/L ).* The measuring method is an adaptation of the USE
2. SPECIFICATIONSCircuit Custom one-chip of microprocessor LSIcircuit.Display LCD size : 41 mm x 34 mmRange Free chlorine ( CL ) :0.00 to 3.50 ppm ( m
Operating 0 to 50 .℃TemperatureOperating Less than 85% R.H.HumidityPower Supply DC 1.5 V battery ( UM4, AAA ) x 6 PCs, or equivalent.Power Curre
3. FRONT PANEL DESCRIPTION3-1 Cover of Testing bottle3-2 Container of Testing bottle3-3 Display3-4 Hold Button ( Esc Button )3-5 TEST/CAL Button3-6 Po
4. MEASURING PROCEDURE4-1 Measurement ConsiderationWhite markon the upposition of" Testing Bottle "White markon the edge of the " Co
3)Insert the " Testing bottle " to the bottom of" Container " ( 3-2, Fig. 1 ) completely.4)Before the measurement, it should envel
3)Insert the " Testing bottle " to the bottom of" Container " ( 3-2, Fig. 1 ) completely.Before the measurement, it should envelop