Sper-scientific 850037 Large Display Conductivity Pen Manual de usuario

Busca en linea o descarga Manual de usuario para Instrumentos de medición Sper-scientific 850037 Large Display Conductivity Pen. Sper Scientific 850037 Large Display Conductivity Pen User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - Conductivity Pens

Large Display Salinity & Conductivity Pens850036 & 850037

Pagina 2

105. Press SET for less than one second to conrm the selection or press SET for more than 2 seconds to return to P3.0 without saving. 6. When in P3

Pagina 3 - TABLE OF CONTENTS

11Range 2 Calibration Information: 4. When in P4.1, press ▲ to enter P4.2 to review the last calibration concentration. If range 2 is not yet calibr

Pagina 4

12To manually select the measurement range:1. Press SET to turn the pen on.2. Press ▲ for more than 2 seconds to select range 1. Press ▲ for more th

Pagina 5

13the pen will retain the calibration data for other ranges that have not yet been re-calibrated.Selecting the Calibration ScheduleCalibration is nece

Pagina 6 - LCD DISPLAY

14Salinity Calibration1. Place the probe into deionized or distilled water for about 30 minutes to rinse the probe.2. Select the NaCl standard for c

Pagina 7 - Salinity Pen

159. When “CAL” stops ashing on the LCD, press SET for less than 1 second to conrm the value. The pen will return to Measurement Mode for salt meas

Pagina 8

16MEASUREMENT PROCEDURES1. Remove the probe cover from the pen. 2. Press SET to turn the pen on. 3. The pen defaults to auto-ranging mode. The au

Pagina 9 - SETUP MODE

175. Rinse the probe tip with de-ionized or distilled water before use to remove any impurities adhering to the probe. If the probe tip is dehydrate

Pagina 10

182. Press HLD again to release the hold function. Turning the Pen Off1. Press SET to turn the pen off from Measurement Mode. 2. To turn the pen o

Pagina 11 - Range 2 10.1 ~ 70.0 ppt

19Accuracy±1% Full Scale± 1 digit (0.0~10.0 ppt)or ±2% Full Scale±1 digit (10.1~70.0 ppt)Range0.00~10.00 ppt (NaCl) or 10.1~70.0 ppt (NaCl)Resolution

Pagina 12

2Large Display Salinity & Conductivity Pens 850036 & 850037Copyright ©2013 by Sper ScienticALL RIGHTS RESERVEDPrinted in the USAThe content

Pagina 13 - Model Concentration

20ERROR CODESSalinity- - - The pen is in manual ranging 1 but the salinity measurement is higher than 10.0 ppt; Press ▲ for more than 2 seconds to c

Pagina 14

21Temperature E01 The temperature circuit is damaged; Send the pen to Sper Scientic for repair. E02 The temperature value is below the range lim

Pagina 16

23Large Display Conductivity Pen 850037

Pagina 17

24SETUP MODEThe advanced Setup Mode allows you to customize the pen’s preferences and defaults: 1. Press SET to turn the pen on. 2. From Measureme

Pagina 18

25Temperature Units (tUt)To select the Celsius or Fahrenheit temperature scale:1. Press SET for 2 seconds to enter setup. 2. “t” will appear on the

Pagina 19 - SPECIFICATIONS

26Reference (Normalization) Temperature (tnr)After saving the temperature unit setting, the pen will automatically enter the reference temperature se

Pagina 20

27Temperature Coefcient (tCo) 1. Press ▲ or ▼ to change the temperature coefcient from 0.0 to 4.0.2. Press SET for less than one second to conrm

Pagina 21

28Resetting the Pen (rSt)When the pen is reset, all parameters (including calibration information) will be reset to factory default values. 1. Press

Pagina 22

29Calibration Review (CAL)1. Press SET for 2 seconds to enter setup. 2. Press ▲ to select the calibration review setting. “CAL” will appear on the p

Pagina 23 - Conductivity Pen

3TABLE OF CONTENTSINTRODUCTION. . . . . . . . . . . . . . 4FEATURES. . . . . . . . . . . . . . . . . 5MATERIALS SUPPLIED

Pagina 24

30P4.1 and P4.2 Modes allow review of the calibration information but are not used during the calibration procedure. When in Temperature Parameter (P1

Pagina 25

31Selecting the Measurement Range (rAn)There are two measurement ranges. The pen defaults to auto-rangingConductivityRange 10 ~ 1999 uSRange 20 ~ 19.

Pagina 26

32CALIBRATIONSelecting the Calibration Standard SolutionFor best results, select a conductivity standard near the sample value that you are measuring

Pagina 27

33Conductivity Measuring RangeRecommended Cal. Solution Range1 0 ~ 1999 uS 600 ~ 1700 uS2 0 ~ 19.99 mS 6.0 ~ 17.00 msThe previous calibration data wil

Pagina 28

34The default value of the reference temperature is set at 25°C. To normalize to a different value, see page 33. Determine the calibration standard va

Pagina 29

35Conductivity Calibration 1. Place the probe into deionized or distilled water for about 30 minutes to rinse the probe.2. Select the conductivity s

Pagina 30

367. Press CAL for more than 2 seconds to begin calibration. The conductivity value of the solution will ash on the LCD.8. Press ▲ or ▼ to adjust t

Pagina 31 - CALIBRATION MODE (CAL)

37Although the values differ less than 30%, the 22 uS is still over the range limit. The maximum input value is 19.99 uS. Example 2: Standard: 1.6 mS;

Pagina 32

38Note…When the pen is switched from Measurement Mode to Calibration Mode, it will display the factory default value. Therefore, if the pen was previo

Pagina 33

394Remove the cover from the bottom of meter to expose the electrode.Electrode 3. The meter is default in auto-rangingstatus once powered

Pagina 34

4INTRODUCTIONThe Sper Scientic Large Display Salinity (Model 850036) and Conductivity (Model 850037) Pens provide highly stable and accurate readings

Pagina 35 - CALIBRATION

405. Set the temperature coefcient to the correct value. The pen defaults to 2.1% per °C, which will typically provide good results. See page 27 to

Pagina 36

419. Stir the probe gently to create a uniform sample. Wait about 15 minutes for the reading to stabilize.10. The unit of measure will ash on the L

Pagina 37

42Accuracy± 1% Full Scale± 1 digitRange0 ~ 1999 uS or 0 ~ 19.99 mSResolution 1 uS or 0.01 mSTemperature Accuracy± 2°F± 1°CTemperature Resolution0.1°C

Pagina 38 - Electrode

43ERROR CODESConductivity- - - The pen is in manual ranging 1 but the conductivity measurement is higher than 1999 uS; Press ▲ for more than 2 secon

Pagina 39

44Temperature E01 The temperature circuit is damaged; Send the pen to Sper Scientic for repair. E02 The temperature value is below the range limi

Pagina 40

45Appendices

Pagina 41

46APPENDIX A:Program Function Default NoteP1.1 Select °C/°F °C Temp unitP1.2 Reference Temperature25°C Selectable 25°C or 20°CP1.3 Temperature Coefci

Pagina 42

47APPENDIX B:Temperature EffectConductivity measurements are temperature dependent; if the temperature increases, conductivity increases. For example,

Pagina 43

48is the value taken at the measurement temperature. The Large Display Salinity (Model 850036) and Conductivity (Model 850037) Pens automatically use

Pagina 44

49Note… The correction is accurate only within a limited temperature range around T1 and T2. The greater the difference between T and Tref, the highe

Pagina 45 - Appendices

5FEATURES• IP65 waterproof housing• Dual display with ATC• Data hold• Pen size for portability • Built-in probe with protective cap• One touch, multi-

Pagina 46 - APPENDIX A:

50Acids: 1.0 - 1.6%/°C Bases: 1.8 - 2.2%/°C Salts: 2.2 - 3.0%/°C Drinking water: 2.0%/°C Ultra-pure water: 5.2%/°C Average temperature coefcients

Pagina 47

51Battery Replacement, Maintenance, & Troubleshooting

Pagina 48

52- 34 - BATTERY REPLACEMENT Note…If the pen will not be in use for a month or longer, remove the batteries and replace when ready to use the pen. BAT

Pagina 49 - APPENDICES

53MAINTENANCEEnsure that the probe is clean before use and storage. To clean:1. Remove the probe cap.2. Rinse the probe carefully in deionized water

Pagina 50

54TROUBLESHOOTINGPen does not turn on:1. Press SET for more than 1 second. 2. Check that the batteries are in place, making good contact, and place

Pagina 51 - & Troubleshooting

5521Error codeFigA F igBFigCFigDtMany air bubbles adhered!To reduce the air bubbles, stir theelectrode completely and better to dip the el

Pagina 52 - BATTERY REPLACEMENT

56WARRANTYSper Scientic warrants this product against defects in materials and workmanship for a period of one (1) year from the date of purchase, an

Pagina 53

6POWER SUPPLYThe pen is powered by 4 LR44 batteries. Check the batteries when:• Using the unit for the rst time• The low battery indicator appears

Pagina 54

7Large Display Salinity Pen 850036

Pagina 55

8SETUP MODEThe advanced Setup Mode allows you to customize the pen’s preferences and defaults: 1. Press SET to turn the pen on. 2. From Measurement M

Pagina 56 - WARRANTY

95. When in P1.0, press ▲ to advance to the next setup function or press SET for more than 2 seconds to return to Measurement Mode. Resetting the Pen

Comentarios a estos manuales

Sin comentarios